System Overview
1-16
Do not slope the table top (the trendelenburg or reverse trendelenburg
position) more than 15
º
when sliding the table top more than 230mm (head
end).
If the table top is already sloped toward the foot end (Reverse
Trendelenburg position), exercise caution when raising the back plate. The
head plate can collide with the patient’s head.
If the table top is already slid toward the head end, be sure to keep the table
top horizontal. Do not slope or tilt the table top under this condition.
In case of excessive tilted or sloped body positions, make sure the patient is
well secured.
Before sliding the table top toward the head end, make sure the table top is
horizontal.
Exercise caution when adjusting the table top or the accessories. Do not
pinch anybody.
When adjusting the table top or the accessories, carefully watch the
positions of the patient, table top, accessories and other things inside the
operating room. Be sure to keep those things from colliding with each other.
Make sure the indoor environment complies with the installation
requirements listed in this manual.
Make sure the ground is level before installing the operating table.
If the integrity of the external protective conductor in the installation or its
arrangement is in doubt, run the operating table from its internal battery.
When the operating table is not in use, make sure it is placed on a level
ground, the castors are properly locked, the table top is in the 0 position and
its height is lower than 940mm.
The operating table can function safely and effectively only if it is
maintained properly and regularly. Be sure to maintain the operating table as
instructed in this manual.
Do not maintain and service the operating table when the patient is lying on
it.
Before maintaining or servicing the operating table, make sure it is
disconnected from the mains supply.
Complete electrical / safety inspections should be conducted every year by
personnel authorized by Nanjing Mindray.
This operating is not suitable for use in the presence of a flammable
anesthetic mixture with air or with oxygen or with nitrous oxide.
Do not use the cleaning agents or disinfectants containing alcohol where
such medical devices as electrosurgical units, defibrillators and defibrillator
monitors may be used. Otherwise, it can cause fire and/or explosion
Содержание HyBase 8300
Страница 1: ...Operator s Manual HyBase 8300 HyBase 8500 Operating Table ...
Страница 2: ......
Страница 84: ...This page is intentionally blank ...
Страница 106: ...This page is intentionally blank ...
Страница 110: ...This page is intentionally blank ...
Страница 122: ...This page is intentionally blank ...
Страница 132: ...This page is intentionally blank ...
Страница 142: ...This page is intentionally blank ...
Страница 150: ...Appendices 10 8 3 Back plate up down up 90 5 down 45 5 ...
Страница 153: ...Appendices 10 11 6 Dual joint head plate up down up 60 5 down 60 5 7 Reverse trendelenburg 20 5 Back plate down 20 5 ...
Страница 158: ...Appendices 10 16 3 Back plate up down up 90 5 down 45 5 ...
Страница 161: ...Appendices 10 19 6 Dual joint head plate up down up 60 5 down 60 5 7 Reverse trendelenburg 20 5 Back plate down 20 5 ...
Страница 167: ...Appendices 10 25 If you want to know more information please see the accessories brochure for details ...
Страница 175: ...This page is intentionally blank ...
Страница 176: ...P N 046 006753 00 2 ...