Daily Operations
2-25
2.3.4.8 Limit Icon
This operating table features a limit indication function. When the table top moves to a limit
position, the corresponding parts of the operating table icon will be yellowed and the override
panel beeps (one tone). See Figure 2-15 for example. This function applies to the following
movements:
Trendelenburg/ reverse trendelenburg position
Tilt to left/right
Back plate up/down
Flex/re-flex position.
Figure 2-15
If the 0-position function is activated, the table top automatically levels and
slides to the centre position. Closely watch the table top movement, and
release the key when the table top no longer moves.
2.3.4.9 Error Icon
If the error icon is displayed, contact the customer service department authorized by Nanjing
Mindray for help.
Содержание HyBase 8300
Страница 1: ...Operator s Manual HyBase 8300 HyBase 8500 Operating Table ...
Страница 2: ......
Страница 84: ...This page is intentionally blank ...
Страница 106: ...This page is intentionally blank ...
Страница 110: ...This page is intentionally blank ...
Страница 122: ...This page is intentionally blank ...
Страница 132: ...This page is intentionally blank ...
Страница 142: ...This page is intentionally blank ...
Страница 150: ...Appendices 10 8 3 Back plate up down up 90 5 down 45 5 ...
Страница 153: ...Appendices 10 11 6 Dual joint head plate up down up 60 5 down 60 5 7 Reverse trendelenburg 20 5 Back plate down 20 5 ...
Страница 158: ...Appendices 10 16 3 Back plate up down up 90 5 down 45 5 ...
Страница 161: ...Appendices 10 19 6 Dual joint head plate up down up 60 5 down 60 5 7 Reverse trendelenburg 20 5 Back plate down 20 5 ...
Страница 167: ...Appendices 10 25 If you want to know more information please see the accessories brochure for details ...
Страница 175: ...This page is intentionally blank ...
Страница 176: ...P N 046 006753 00 2 ...