System Overview
1-15
operating table, when a patient lying on the operating table.
The maximum load of this operating table is 460kg. However, in case of
certain special conditions, the load capacity is further limited and see
chapter A.3 for details. Be sure to take those conditions into consideration
when using the operating table. If you have any questions, contact the
customer service department authorized by Nanjing Mindray or your local
distributor.
If accessories are used, make sure the weights of the accessories installed
on the operating table and of the patient add up no more than the
permissible overall load.
To avoid infection, be sure to cover the table-top pad with a sterile surgical
drape before transferring the patient.
Do not use wet surgical drapes on the operating table.
Transferring the patient inappropriately can tip over the operating table. Be
sure to transfer the patient from either side of the operating table. Do not
transfer the patient from either the head or the foot end.
Before transferring the patient, make sure the operating table is placed on a
level ground, the castors are properly locked and the table top is in the best
balanced state (0 position: table top horizontal and slid to the central
position). Do not transfer the patient otherwise.
Placing the patient improperly can tip over the operating table. Be sure to
place the patient in the right direction. Do not place the patient’s trunk on
the leg plates.
When placing the patient, make sure the patient’s center of gravity is as
close to the column as possible. Otherwise, the operating table can tip over.
Improper body positions can damage the patient, especially in case of
excessive tilted or sloped position. When using this operating table, ensure
proper body positions and regularly check the position during the whole
surgery.
To avoid damaging the patient’s respiratory system, nervous system or
circulating system, make sure the patient is properly positioned and keep a
close eye on the patient during the whole surgery.
When using the operating table, be sure to press the exact function key
matching the patient’s body position. Otherwise, the patient can be
damaged.
Before positioning the patient, make sure the castors are properly locked.
The speed of table-top movement can change depending on the patient’s
weight. Closely watch the table top during adjustment.
Do not slide the table top when sloping the table top (the trendelenburg or
reverse trendelenburg position) more than 15
º
.
Содержание HyBase 8300
Страница 1: ...Operator s Manual HyBase 8300 HyBase 8500 Operating Table ...
Страница 2: ......
Страница 84: ...This page is intentionally blank ...
Страница 106: ...This page is intentionally blank ...
Страница 110: ...This page is intentionally blank ...
Страница 122: ...This page is intentionally blank ...
Страница 132: ...This page is intentionally blank ...
Страница 142: ...This page is intentionally blank ...
Страница 150: ...Appendices 10 8 3 Back plate up down up 90 5 down 45 5 ...
Страница 153: ...Appendices 10 11 6 Dual joint head plate up down up 60 5 down 60 5 7 Reverse trendelenburg 20 5 Back plate down 20 5 ...
Страница 158: ...Appendices 10 16 3 Back plate up down up 90 5 down 45 5 ...
Страница 161: ...Appendices 10 19 6 Dual joint head plate up down up 60 5 down 60 5 7 Reverse trendelenburg 20 5 Back plate down 20 5 ...
Страница 167: ...Appendices 10 25 If you want to know more information please see the accessories brochure for details ...
Страница 175: ...This page is intentionally blank ...
Страница 176: ...P N 046 006753 00 2 ...