Troubleshooting
6-2
Error
Cause
Solution
The operating table cannot
be turned on.
1. The battery is dead and the
mains supply is not
connected.
2. The power switch on the
base of the operating table is
off.
3. The operating table is in
transport mode.
1. Check whether the mains
supply is connected.
2. Check whether the battery
is sufficiently charged.
3. Check whether the power
switch is on.
4. Press and hold the
key on the override panel for
3 seconds.
When you press the
key, the table top can not be
slid and a sound signal (one
tone) is heard.
The table top is in the
extended range.
Press and hold
+
.
The battery indicator is
blinking.
The battery is low.
Connect the operating table
to the mains supply.
Содержание HyBase 8300
Страница 1: ...Operator s Manual HyBase 8300 HyBase 8500 Operating Table ...
Страница 2: ......
Страница 84: ...This page is intentionally blank ...
Страница 106: ...This page is intentionally blank ...
Страница 110: ...This page is intentionally blank ...
Страница 122: ...This page is intentionally blank ...
Страница 132: ...This page is intentionally blank ...
Страница 142: ...This page is intentionally blank ...
Страница 150: ...Appendices 10 8 3 Back plate up down up 90 5 down 45 5 ...
Страница 153: ...Appendices 10 11 6 Dual joint head plate up down up 60 5 down 60 5 7 Reverse trendelenburg 20 5 Back plate down 20 5 ...
Страница 158: ...Appendices 10 16 3 Back plate up down up 90 5 down 45 5 ...
Страница 161: ...Appendices 10 19 6 Dual joint head plate up down up 60 5 down 60 5 7 Reverse trendelenburg 20 5 Back plate down 20 5 ...
Страница 167: ...Appendices 10 25 If you want to know more information please see the accessories brochure for details ...
Страница 175: ...This page is intentionally blank ...
Страница 176: ...P N 046 006753 00 2 ...