C - 4
Operator’s Manual
C Electrical Safety Inspection
C.8
Mains on Applied Part Leakage
The Mains on Applied Part test applies a test voltage, which is 110% of the mains voltage using a
limiting resistance, to selected applied part terminals. Current measurements are then taken
between the selected applied part and earth. Measurements are taken with the test voltage (110% of
mains) on applied parts in the normal and reverse polarity conditions.
The following outlet conditions apply when performing the Mains on Applied Part test.
•
Normal Polarity;
•
Reversed Polarity.
LIMITS
For BF
applied parts: 5000
μ
A.
C.9
Patient Auxiliary Current
Patient Auxiliary currents are measured between any selected Applied Part connector and the
remaining Applied Part connectors. All measurements may have a true RMS response.
The following outlet conditions apply when performing the Patient Auxiliary Current test.
•
normal polarity (Normal Condition);
•
reverse polarity (Normal Condition);
•
normal polarity with open neutral (Single Fault Condition);
•
reverse polarity with open neutral (Single Fault Condition);
•
normal polarity with open earth (Single Fault Condition);
•
reverse polarity with open earth (Single Fault Condition).
LIMITS
For BF
applied parts,
•
100
μ
A in Normal Condition.
•
500
μ
A in Single Fault Condition.
Содержание Anesus ME7T
Страница 2: ......
Страница 58: ...This page intentionally left blank ...
Страница 154: ...This page intentionally left blank ...
Страница 164: ...This page intentionally left blank ...
Страница 182: ...This page intentionally left blank ...
Страница 190: ...This page intentionally left blank ...
Страница 208: ...This page intentionally left blank ...
Страница 254: ...This page intentionally left blank ...
Страница 264: ...This page intentionally left blank ...
Страница 280: ...This page intentionally left blank ...
Страница 311: ......
Страница 312: ...P N 046 018839 00 5 0 ...