6 Image Acquisition
Operator’s Manual
6 - 9
6.6
Anatomical M Mode
For an image in the traditional M mode, the M-mark line goes along the beams transmitted from the
probe. Thus it is difficult to obtain a good plane for difficult-to-image patients who cannot be
moved easily. However, in the Anatomical M mode, you can manipulate the M-mark line and move
it to any position at desired angles. The system supports anatomical M scanning (including Free
Xros M mode and Free Xros CM mode) in 2D imaging modes (B, Color, Power and TVI mode).
CAUTION
Anatomical M Mode and Color Anatomical M mode images are provided for
reference only, not for confirming diagnoses. Compare the image with those of
other machines, or make diagnoses using non-ultrasound methods.
TIP:
Free Xros M imaging and Free Xros CM imaging are options.
6.6.1 Linear Anatomical M (Free Xros M)
Free Xros M imaging is supported on frozen B image, B+M image and B+Power/Color/TVI image.
Perform the following procedure:
1.
Adjust the probe and image to obtain the desired plane in real-time B mode or M mode.
Or select the B mode cine file to be observed.
2.
Tap [Free Xros M] on the touch screen to enter Free Xros M mode or press the user-defined
key to enter Free Xros M mode.
There are 3 M-mark lines available, each with a symbol of “A”, “B” or “C” at one end as
identification.
3.
Adjust the sampling line (single line or couple of lines) to obtain optimized images and
necessary information.
–
Tap [Show A], [Show B] or [Show C] on the touch screen to adjust the sampling line. The
corresponding sampling line and the Free Xros M image appear on the screen. Then,
activate the sampling line.
–
Tap [Display Cur.] or [Display All] on the touch screen to select whether to display the
image of the current M-mark line or all.
You can choose to display the sampling line on the current image or all.
–
Press <Set> to switch among the sampling lines and press <Cursor> to show the cursor.
4.
Adjust the image parameters to obtain optimized images.
5.
Press <B> to return to real-time B mode.
6.6.2 Free Xros CM (Curved Anatomical M-Mode)
In Free Xros CM mode, the distance/time curve is generated from the sample line manually
depicted anywhere on the image. Free Xros CM is used for TVI and TEI modes.
Содержание Anesus ME7T
Страница 2: ......
Страница 58: ...This page intentionally left blank ...
Страница 154: ...This page intentionally left blank ...
Страница 164: ...This page intentionally left blank ...
Страница 182: ...This page intentionally left blank ...
Страница 190: ...This page intentionally left blank ...
Страница 208: ...This page intentionally left blank ...
Страница 254: ...This page intentionally left blank ...
Страница 264: ...This page intentionally left blank ...
Страница 280: ...This page intentionally left blank ...
Страница 311: ......
Страница 312: ...P N 046 018839 00 5 0 ...