P377
Initial Zero
Resets the initial zero.
The initial zero is the reference zero to which all subsequent operator initiated zero
calibrations are compared in determining whether they have deviated beyond the
zero limit (P370).
(
f
=1)
Note:
Refer to Initial Zero on page 46 for execution.
P388
Initial Span
Resets the initial span for the span selected, multispan 1 to 8 (P388-01 to –08).
The initial span is the reference to which all subsequent span calibrations are
compared in determining whether they have deviated beyond an accumulated
±
12.5% of the initial span.
(
f
=1)
Note:
Refer to Initial Span on page 49 for execution.