Scale – LVDT
*Where separation between the BW500 and LVDT conditioner exceeds 150 m (500 ft):
•
remove the jumpers BW500 terminal 11/12 and 13/14
•
run additional conductors from :
BW500 terminal 12 to integrator terminal block ‘+EXC’
BW500 terminal 13 to integrator terminal block ‘-EXC’
For further connection information on specific LVDT’s consult Milltronics.
♦
Shields are common, but not grounded to chassis. Run cable shields through SHLD terminals
and ground at BW500 only.
Milltronics LVDT Conditioner
belt scale
with LVDT
Milltronics BW500
*
maximum cable run LVDT to Conditioner3 m (10 ft)