Maintenance
Safety Precautions
•
Never use solvents or abrasives that could damage the equipment’s surfaces.
•
Do not put liquids directly in contact with the internal parts of the pump or the
equipment’s electrical components.
•
Do not turn the pump upside down.
Ongoing Maintenance
•
Clean the pump and accessories after each use. See
.
NOTE
It is important to autoclave the drain tray components and the drain tubing
after each use.
•
Perform the pressure sensor verification every six months. See
•
The Steritest
®
Symbio Pump should be serviced by a certified technician
annually. This servicing includes a check of the operating performances
and calibration of the pressure sensors. Contact our sales representative or
Technical Service for more information about the service agreements.
Preparing the Pump for Shipment
Follow these steps to prepare the pump for shipping for servicing or any other
reason.
1.
Press the
button to switch the pump on.
2.
Enter the Standard Mode.
3.
Press the
button to close the pump head.
4.
With the pump head in its closed position, press and hold the
button on
the control panel for a few seconds to switch off the pump.
NOTE
This allows the pump head to remain closed when the pump is off.
5.
Switch off the communication hub by pressing the on/off switch (position Ο).
Disconnect it from power source.
6.
Disconnect the communication hub pump connection cord from the pump
power inlet.