C
heck the tubing and do one of the following:
•
If the tubing is damaged:
•
Turn and press the control knob to select Yes. The following screen
displays:
•
Discard the Steritest
®
device and unpack a new one. Press the
button.
•
Restart the pressure sensor test procedure as explained in
the Pump, Pressure Sensor Test
•
If the tubing is not damaged,
•
Turn and press the control knob to select No. The following screen
displays:
•
Re-install the tubing in the pump head.
•
Press the
button. The screen for the selection of the pressure
sensor to verify displays with the last sensor tested selected.
•
Press the control knob to confirm and restart the procedure from
step 7 (step 5 on the pump screen) or step 15 (step 13 on the pump
screen) of the
procedure.