Re-install the tubing in the pump head.
•
Press the
button. The selection screen for the pressure sensor to
verify displays with the last sensor tested selected.
•
Press the control knob to confirm and restart the procedure from
step 7 (step 5 on the pump screen) or step 15 (step 13 on the pump
screen)
If the pump must be stopped manually again during the test, the following
message displays. The verification procedure cannot be completed.
If this happens:
•
Remove the pressure control kit for Steritest
®
Pumps from the tested
canister.
•
Remove the bottle from the bottle holder and place it upright on the
workbench.
•
Open the tubing clamp.
•
Press the
button to open the pump head and return to the
Configuration
menu.
•
Contact Technical Service.
NOTE
Until the pressure sensors issue has been resolved, the pressure level in
the canisters should be assessed exclusively through the use of the red
plugs.