NOTE
To stop the pump before the end of the timer value, press the control knob.
A message asking if filtration should resume with remaining time displays.
To stop the pump, turn and press the control knob to select
No
.
Using the Waste Overfill Sensor in Standard Mode
In Standard Mode, when the maximum liquid waste level is reached, the waste
overfill sensor icon changes from blue to red. An alert beep sounds from the
pump. An orange LED lights up on the waste overfill sensor. The pump continues
to run.
When the pump is stopped manually, the overfill alert screen displays:
The red waste overfill sensor icon and the warning symbol remain on the screen
until the waste receptacle is emptied or replaced.
NOTE
The test in progress can be finished before the waste container is emptied
or replaced.
Once the waste receptacle is emptied or replaced, the warning symbol is no
longer displayed on the Standard Mode screen, the waste overfill sensor icon
returns to blue, and the orange LED no longer lights up on the waste overfill
sensor.
Using the Pump in Test Method Mode
The
Test Method Mode
is available only if test methods have been loaded in the
pump. Test methods are created and transferred to the Steritest
®
Symbio Pump
using the Steritest
®
Symbio Software. The software and software user guide are
available at
www.millipore.com/steritest-software
.
To use the pump in
Test Method Mode
:
1.
Access the main screen.