Recommended Cleaning Agents
The compatibility of the following cleaning agents with the Steritest
®
Symbio
Pump has been verified.
Cleaning Product
Validated
dilution
1
Ethanol (<30%)
Didecyldimethylammonium chloride (<1%)
Propanol (<1%)
N-(3-aminopropyl)-N-dodecylpropane-1,3-diamine
(<1,5%)
Ready to use
2
Isopropyl alcohol 70%
Ready to use
3
Ethanol 70%
Ready to use
4
Sodium hypochlorite
250 ppm
5
Peracetic acid at (5 to 15%)
Acetic acid at (5 to 15%)
Hydrogen peroxide (15 to 30%)
2.5%
6
Ethanol (25 to 50%)
Polyhexamethylene biguanide hydrochloride (<2,5%)
N-(3-aminopropyl)-N-dodecylpropane-1,3-diamine
(<2,5%)
Didecyldimethylammonium chloride (<1%)
Ready to use
Disassembling the Pump
The pump must be disassembled for cleaning.
1.
Ensure that the pump head is open.
2.
Press the
to shut down the equipment. The cover disengages from the
automatic closure system.
3.
When shutdown is complete, switch off the communication hub using the on/
off switch (position Ο) and disconnect the hub from the power source.