Symbio Waste Overfill Sensor for Solid Containers
This accessory prevents the liquid waste container from overflowing, and is
especially helpful when the waste container is placed out of the operator’s view.
When the liquid reaches the maximum filling level, an alert appears on the
Steritest
®
Symbio Pump display (see
Using the Waste Overfill Sensor in Standard
Using the Waste Overfill Sensor in Test Method Mode
).
NOTE
The test in progress can be finished before the waste container is emptied
or replaced.
To use the overfill sensor:
1.
Place the sensor around the waste container. Use the belt to secure the
position.
NOTE
Keep enough space at the top of the container so that the filtration in
progress can be completed before overflowing occurs.
2.
Connect the waste overfill sensor connector to the dedicated port of the
communication hub.
NOTE
The waste overfill sensor can be connected at any time, regardless of pump
status.
3.
If the pump is off, switch on the communication hub by pressing the on/
off switch (position I). The Welcome screen displays. After the autotest
sequence, the Accessories and data sources screen displays. The check in
the check box next to Overfilling sensor indicates that the waste overfill
sensor is activated.