Two operating modes are available:
Footswitch Mode
Symbol
Start/Stop
Press the footswitch once to start the pump. Press again to stop the
pump. Press it a third time to restart the pump.
If the timer is activated (the timer value is set to anything but off)
and the footswitch is pressed to stop the pump before the end of the
set value, a confirmation screen asks if the filtration should resume
with remaining time or if the timer value should be canceled and the
pump stopped.
Continuous
The pump runs as long as the foot remains on the footswitch.
If the timer is activated (the timer value is set to anything but
off) and the footswitch is released before the timer reaches the
end of the set value, the pump continues to run until the set value
is reached, even if the footswitch is pressed again. To stop the
pump before the end of the timer value, press the control knob.
A confirmation screen asks if the filtration should resume with the
remaining time or if the timer value should be canceled and the
pump stopped.
2.
Turn and press the control knob to select the desired footswitch operating
mode.
NOTE
The enabled operating mode is indicated by an icon at the top left of the
main screen and use screen.
3.
Press the
button to return to the
Configuration
menu.
Syringe Support
The syringe support is an optional accessory (see
Overview of Optional Accessories
. Use the syringe support menu to activate
or deactivate the syringe support.
To activate or deactivate the syringe support:
1.
In the
Configuration
menu, turn and press the control knob to select
Syringe Support.
If the syringe support is not connected to the communication hub, the
following screen displays (to connect the syringe support, see