No Pressure Increase Detected
During the pressure sensors test, if no pressure increase is detected after
about 30 seconds in the tested canister, the test stops and the following
screen displays:
Remove the bottle from the bottle holder and place it upright on the
workbench.
Ensure that the tubing of the tested canister is not clamped at the bottle end.
1.
Press the
button. The following screen displays:
Ensure that the pressure gauge is secured on the canister and that the
yellow plugs are on the canister outlets.
2.
Press the
button. The following screen displays: