NOTE
On the LFH version, a intermediate position exists for use when the pump
must be carried without removing the drain tray.
4.
Connect the silicone drain tubing to the drain tray.
5.
Place the other end of the drain tubing directly into a waste container (for
LFH version) or connect it to the isolator table drain (for ISL and FLEX
versions).
NOTE
To prevent back flow of fluid and reverse pressure into the drain tray,
ensure that the tubing does not touch the liquid in the waste container and
that the waste container is open or correctly vented. Also ensure that the
drain tray is higher than the waste container, and that the tubing between
them is straight, without any bends or loops.
A waste overfill sensor can be used to automatically check the liquid level in
the waste container (see
Overview of Optional Accessories
).
Follow safety procedures for the handling of the filtrated liquid.
Dispose of the liquid in accordance with local regulations.
Drain tubing