23.
P
ress the control knob to stop the pump when the canisters are empty. The
following screen displays:
Remove the red plugs from the canister vents.
Press the
button to open the pump head.
Discard the Steritest
®
device.
24.
Press the
button to return to the configuration menu.
Pressure Sensor Test Issues
Overpressure
1.
During the pressure sensors test, if the pressure in the canister exceeds 53
psi, manually stop the pump by pressing the control knob. The following
screen displays:
Remove the pressure control kit for Steritest
®
Pumps from the tested
canister.
Remove the bottle from the bottle holder and place it upright on the
workbench.
Open the tubing clamp.
2.
Press the
button.