If the same issue occurs again during the test, the following message displays,
and the verification procedure cannot be completed:
If this happens :
•
Remove the pressure control kit for Steritest
®
Pumps from the tested
canister.
•
Remove the bottle from the bottle holder and place it upright on the
workbench.
•
Open the tubing clamp.
•
Press the
button to open the pump head and return to the
Configuration menu.
•
Contact Technical Service.
NOTE
Until the pressure sensors issue has been resolved, the pressure level in
the canisters should be assessed exclusively through the use of the red
plugs.
Pressure Outside the Acceptable Range
During the pressure sensors test, if the pressure reading on the pressure gauge
when the pump stops is below 37 psi or higher than 53 psi, do the following:
•
Ensure that the pressure control kit for Steritest
®
Pumps is calibrated.
•
Repeat the test. If the problem persists, contact technical service and have a
certified technician calibrate the pressure sensors.