Reassembling the Pump
T
o reassemble the pump:
1.
Ensure that the communication hub is disconnected from the power source.
2.
Reattach the control knob by pushing it onto the post.
3.
If the pump head rotor has been removed, reinstall it by aligning the notch
in the rotor with the guide part.
4.
Insert the pump head rotor mounting screw and tighten using a 4mm hex
wrench.
5.
Install the protective cover on the pump head:
•
Position the cover over the pump head and engage the clamp rotation spindle with
its housing in the cover.
Notch
Guide part
SB04a