7.
Clean the control knob with a nonwoven wipe moistened with one of the
recommended cleaning agents listed in this section. Rinse with clean water.
Optional: Autoclave the control knob at 121 °C for 30 minutes or 134 °C for
10 minutes.
8.
Clean the pump head with a nonwoven wipe moistened with one of the
recommended cleaning agents listed in this section.
9.
Clean the drain tray support and drain tray container with a nonwoven
wipe moistened with one of the recommended cleaning agents listed in this
section. Rinse with clean water.
Completely disassemble the drain tray components. Autoclave the drain
tray support, the drain tray container, and the drain tubing for 121 °C for 30
minutes or 134 °C for 10 minutes.
NOTE
Perform this operation daily. Spare drain trays are available (see
).
10.
Clean the bottle holder parts using a nonwoven wipe moistened with one
of the recommended cleaning agents listed in this section. Rinse with clean
water.
Optional: Autoclave the bottle holder parts (disassembled) at 121 °C for 30
minutes or 134 °C for 10 minutes.
11.
Clean the optional Steritest
®
holder for sterile bags and the Steridilutor
®
vent
chamber using a nonwoven wipe moistened with one of the recommended
cleaning agents listed in this section. Rinse with clean water.
12.
Clean the optional Steritest
®
carrying tray and Steritest
®
rack using a
nonwoven wipe moistened with one of the recommended cleaning agents
listed in this section. Rinse with clean water.
Optional: Autoclave the tray and the rack at 121 °C for 30 minutes or 134 °C
for 10 minutes.
NOTE
Refer to the ampoule breaker and syringe support user guides for cleaning
recommendations.