14-25
Utilities
Tube/Cap Learn
The
<@LCAP>
command activates the Tube/Cap Learn function. When new profiles are
found, Tube/Cap Learn stores them in the
Tube/Cap Library
.
When
<@LCAP>
is sent from
Terminal
in
ESP
, one of the following outputs will result:
Learning Tube and Cap...
Added to library #xx
Learning Tube and Cap...
Already in library #xx
Learning Tube and Cap...
Unable to locate a cap
Learning Tube and Cap...
Unable to locate a tube
Learning Tube and Cap...
Unable to add (Library full)!
Learning Tube and Cap...
Please enable a cap detecting operation!
Important:
Upon initiation of a Tube/Cap Learn operation, the process searches active
operations (Multi-Operations 1 – 3,
<K260>
) and locates the first operation that has Tube/Cap
Detection enabled. This operation is used to “learn” the tube or cap. If no operation has
Tube/Cap Detection enabled, the following message will be output:
“Please enable a cap detecting operation!”
Содержание EZ Match
Страница 1: ...EZ Match Imager User s Manual P N 83 000044 Rev C...
Страница 12: ...xii EZ Match User s Manual Statement of RoHS Compliance...
Страница 44: ...2 18 EZ Match User s Manual Send Receive Options...
Страница 104: ...4 36 EZ Match User s Manual Postamble...
Страница 122: ...5 18 EZ Match User s Manual Additional Notes about Calibration...
Страница 244: ...8 60 EZ Match User s Manual EZ Trax Output...
Страница 258: ...9 14 EZ Match User s Manual Microscan Symbol Quality Output by ESP...
Страница 270: ...10 12 EZ Match User s Manual New Master Pin...
Страница 370: ...15 14 EZ Match User s Manual Ordered Output Filter...