EZ Match User’s Manual
6-21
Read Cycle
Triggered Capture
Usage:
Useful in applications where each decode must be treated as a discrete
event, regardless of symbol data.
Definition:
The first trigger event starts the read cycle, and subsequent triggers will
continue until the predetermined
Number of Captures
is met, or until the
predetermined
End of Read Cycle
condition is met-- whichever occurs
first.
Note:
If
End of Read Cycle
is set for
New Trigger
and the read cycle
qualifications have not been met, the read cycle will only end
once it
receives the first trigger after reaching the predetermined Number of
Captures setting
.
Processing
Read cycle ends
on New Trigger
or Timeout, as
configured.
Captures on every trigger
Capture
First trigger
starts the read
cycle