Gemini VII 2390 Confirm
Cleaning and Labeling Sample Tubes
Feb 09 Gemini VII 2390 Confirm Installation Instructions and Checklist
53
Cleaning and Labeling Sample Tubes
Sample tubes must be clean and dry before samples are added and weighed. The following
procedures are recommended. Refer to following table for a list of materials needed to clean
and weigh samples.
Materials Required to Clean and Weigh Sample Tubes
1. Turn on the drying oven used for heating the sample tubes and set the temperature to
110 ºC.
2. Check the reservoir of the ultrasonic cleaning unit to make sure it is clean.
Supplied by Micromeritics
Supplied by User
Sample tube
Drying oven
Sample tube brush
Ultrasonic cleaning unit
Stopper for sample tube
Detergent
Sample tube rack
Rubber gloves or lint-free cloth
Sample weighing support
Acetone or isopropyl alcohol
Safety glasses
Waste container
Analytical balance
Pipe cleaners