Calibrating Transducers and Servo Valves
Gemini VII 2390 Confirm
48 Gemini VII 2390 Confirm Installation Instructions and Checklist
Feb 09
Part 4. Operational Verification
Calibrating Transducers and Servo Valves
Calibrate the 1000-Torr Pressure Transducers
1. Ensure that nitrogen is attached to the
Adsorptive
gas inlet port on the instrument and the
pressure is set to 15 psig.
2. Pressurize the ports.
a. Select
Unit [n] > Show Instrument Schematic
,
Unit [n] > Enable Manual Control
.
b. Close all valves (double-click a valve to open or close it).
c. Open valves 4-b and 4-s. Open valve 2 (adsorptive) for 15 seconds.
d. Close all valves.
Verifying the 1000-Torr transducers’ accuracy requires the use of:
• 004-09814-00 Paroscientific Pressure Calibration Standard OR
• 004-09603-00 Crystal Pressure Gauge (if 004-09814-00 is
unavailable)
• 003-09623-01 Vacuum Calibration Standard. (Yellow Jacket)