Calibrating Transducers and Servo Valves
Gemini VII 2390 Confirm
50 Gemini VII 2390 Confirm Installation Instructions and Checklist
Feb 09
e. Press
Start
to begin the zero pressure transducers routine.
f. When the vacuum gauge reaches 50 microns, click
Enter
to zero the pressure
transducers. There is a 3-minute evacuation, then a message that asks if you want to
save the calibration. Generally, you should select
No
.
6. Scale the pressure transducers.
a. Pressurize ports to 780 mmHg as described in step 2 above.
b. Remove the vacuum gauge and install the pressure gauge.
c. Connect the RS232 cable for data transfer.
d. From the main menu, for the unit number being operated, select
Calibration >
Pressure Scale
.The following dialog is displayed.
e. Select the
Sample Port Transducers
and
Po Port Transducers
, if present.