Gemini VII 2390 Confirm
Performing a Reference Analysis
Feb 09 Gemini VII 2390 Confirm Installation Instructions and Checklist
67
c. Edit the description in the
Sample
field. For example, you may wish to add the date.
d. Enter the sample mass you recorded after degassing the sample in the
Sample Mass
field.
e. Click
Save
, the Save as Sample Information File dialog is displayed.
f. Enter a comment in the
Comment
field, then click
Save
to save the file and return to
the Sample Information dialog.
g. Click
Close
to close the dialog.
2.
Select
Unit > Start Analysis
; the Analysis dialog is displayed.
3.
Click
Browse
and select the file you created for this analysis, then click
Open
to return
to the Analysis dialog.
a. Enter the sample’s mass in the
Sample Mass
field.
b. Click
Report After Analysis
. Ensure that
Report after analysis
is selected and that
Be sure the sample compartment door is closed before beginning an analysis. If an
abnormal condition causes the analyzer to operate at an excessive pressure, the
sample or balance tube could dislodge from its port, possibly breaking the Dewar and
cause personal injury or damage to the equipment.