Gemini VII 2390 Confirm
Calibrating Transducers and Servo Valves
Feb 09 Gemini VII 2390 Confirm Installation Instructions and Checklist
49
e. Set the servo target to increase to 780 mmHg and turn the servo on.
f. Turn the servo off when sample pressure stabilizes.
g. Select
Unit [n] > Enable Manual Control
to disable manual control (remove check
mark).
3. Install the vacuum gauge on the sample port.
4. Ensure that a blank tube or plug is installed in the balance port (refer to
).
5. Zero the pressure transducers.
a. Select
Options > Service Test Mode
.
b. Enter the password, then click
OK
.
c. From the main menu, select
Unit[n]
>
Calibration
>
Zero Pressure
. The following
dialog is displayed.
d. Select all transducers.