Installing the Vacuum Pump
Gemini VII 2390 Confirm
8 Gemini VII 2390 Confirm Installation Instructions and Checklist
Feb 09
Preparing the Alumina Trap
1. Insert an O-ring into each of the end fittings.
2. Screw one of the end fittings onto the trap body clockwise.
3. Be sure the trap body is dry and the alumina pellets are cool. Pour the pellets into the trap
until they are level with the top of the trap body.
4. Screw the other end fitting onto the trap and tighten securely by hand.
Be sure alumina pellets are cool. Adding hot pellets to the trap body may warp the
body and cause an improper seal of the O-ring.