Troubleshooting Guide
15 - 27
SCC: PIO3 INOPERATIVE
Input/Output port 3 (PIO3) on the Servo Control Card is not function-
ing properly. Replace the Servo Control Card.
SCC: PSI DAC FAILURE
Pressure limit DAC output on the Servo Control Card is not within
allowable limits. Remove and reapply power, then attempt to operate
the unit again. If the condition persists, replace the Servo Control
Card.
SELECT HEAD: 60/150=1, 150/200=2
Displayed in response to RECALL, 65, after the service access pass-
word is properly entered. May also be displayed if unit memory is lost.
Enter the corresponding number of the injector head to be used with
the system.
SELECT UTILITY FUNCTION
Displayed if unit is powered up with SP1 on the CPU set for the utility
mode. Select the Utility Recall function desired.
SELF TEST INDICATES 60 ML TURRET
Displayed only when the turret is in the 60 ml syringe position.
RECALL, 65 has been accessed in order to change head configura-
tion to the 150/200 injector head. The system has detected that a 60/
150 head is installed, therefore, the 150/200 configuration cannot be
used.
SELF TEST IN PROGRESS VER MV XXX
The unit is running a self-test which occurs at power-up. Wait approxi-
mately 3-10 seconds for completion of the self-test, then continue with
normal operations.
SET LEVEL TO 1 TO CHANGE SCALE
An attempt was made to change the Flow scale at an intermediate
level in a multiple-level injection. Go to level 1, then change the Flow
Scale
SET VOLUME IS TOO HIGH XXX.X REMAINS
Insufficient Volume remains in the syringe for an injection of the
selected Volume. Enter a lower Volume setting or change syringes,
replacing with a full syringe.
Содержание ProVis Mark V
Страница 1: ...Service Manual KMP 870 98411 T 205 Rev D...
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 15: ...Operating Guide 3 1 3 Operating Guide...
Страница 16: ...Mark V ProVis Injection System 3 2...
Страница 30: ...Mark V ProVis Injection System 7 4 Figure 7 1 Block Diagram Servo Control Card...
Страница 31: ...Servo Control Card 7 5...
Страница 37: ...Mark V ProVis Injection System 9 4 Figure 9 1 Block Diagram Control Panel Digital Logic...
Страница 39: ...Mark V ProVis Injection System 9 6 Figure 9 2 Block Diagram Control Panel DC to DC Converter...
Страница 49: ...Mark V ProVis Injection System 10 10 NOTES...
Страница 60: ...Injector Head 13 7 Figure 13 1 Block Diagram Head Card...
Страница 61: ...Mark V ProVis Injection System 13 8 NOTES...
Страница 64: ...Imaging System Interface 14 3 Figure 14 1 Connector Locations...
Страница 67: ...Mark V ProVis Injection System 14 6 Figure 14 2 Block Diagram Imaging System Interface...
Страница 71: ...Mark V ProVis Injection System 14 10 NOTES...
Страница 107: ...Mark V ProVis Injection System 15 36 NOTES...