Troubleshooting Guide
15 - 29
TOTAL VOLUME EXCEEDED
The total volume of a multi-level injection exceeds the maximum pos-
sible volume of the selected syringe size. Change the settings to a
lower total Volume, or, select a larger syringe size.
TOTAL VOLUME OF XXX.X DESIRED?
For Multi-Level injections only. Asks for verification that the operator is
aware of the total Volume selected for all levels of the injection.
“XXX.X” indicates the total injection Volume. Verify that the Volume
displayed is the Volume desired, then press YES.
UNIT DISARMED XX
While the unit was armed, a disarm condition occurred. This message
is displayed for three seconds. The following is the disarm code refer-
ence list which identifies the source of the disarm condition. (For
example; if a panel key on the main keyboard is pressed while the unit
is armed or injecting, the following message would be displayed:
UNIT DISARMED 01
01 Main keyboard
03 Remote panel keyboard
04 FWD/REV load switch
05 Turret
06 Watch Dog failure while arming
07 0300 only - Injector or Aux start detected while arming or injecting
08 0300 only - X-Ray or ISI released while injecting
09 0300 only - X-Ray or ISI detected during 0300 handswitch start
10 0300 only - Handswitch released (armed or injecting)
11 Remote handswitch released
12 Communication link failure
13 PIO3 Interrupt - type undetermined
14 External Disarm
15 PIO2 or PIO3 Interrupt - type undetermined
16 Start switch debounce error
VACANT INJECTION NUMBER
No program exists at the selected PPI storage number.
VERIFY DURATION
An attempt was made to arm the injector without a Duration value
entered or calculated. Enter a Duration value, or, reenter the Flow
Rate and Volume parameters.
VERIFY FLOW RATE
An attempt was made to arm the injector without a Flow Rate value
entered or calculated. Enter a Flow Rate value, or, reenter the Volume
and Injection Duration parameters.
Содержание ProVis Mark V
Страница 1: ...Service Manual KMP 870 98411 T 205 Rev D...
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 15: ...Operating Guide 3 1 3 Operating Guide...
Страница 16: ...Mark V ProVis Injection System 3 2...
Страница 30: ...Mark V ProVis Injection System 7 4 Figure 7 1 Block Diagram Servo Control Card...
Страница 31: ...Servo Control Card 7 5...
Страница 37: ...Mark V ProVis Injection System 9 4 Figure 9 1 Block Diagram Control Panel Digital Logic...
Страница 39: ...Mark V ProVis Injection System 9 6 Figure 9 2 Block Diagram Control Panel DC to DC Converter...
Страница 49: ...Mark V ProVis Injection System 10 10 NOTES...
Страница 60: ...Injector Head 13 7 Figure 13 1 Block Diagram Head Card...
Страница 61: ...Mark V ProVis Injection System 13 8 NOTES...
Страница 64: ...Imaging System Interface 14 3 Figure 14 1 Connector Locations...
Страница 67: ...Mark V ProVis Injection System 14 6 Figure 14 2 Block Diagram Imaging System Interface...
Страница 71: ...Mark V ProVis Injection System 14 10 NOTES...
Страница 107: ...Mark V ProVis Injection System 15 36 NOTES...