Mark V ProVis Injection System
15 - 4
Test Point / Description
Typical Values
10 BMSPOSCMD
Measure the same as test point 6
Buffered M/S Position CMD
when the Mechanical Stop is not
moving.
11 10V REF.
+10.0 +0.1 VDC.
+10 VDC Reference
12 DIG. VELOCITY
+0.1 VDC/ml (+4.0 VDC at 40 ml/sec).
Digital Velocity
13 ANA. VELOCITY
+0.15 V/ml (+6.0 VDC at 40 ml/sec).
Analog Velocity
14 SAFRELO
+24 VDC when disarmed and not
Safe Relay Low
loading; less than +2 VDC when
armed, injecting, or loading.
15 INJ. START
+5 VDC when start switch open; 0 to
Inject Start
+0.1 VDC when start switch closed.
16 INJECTING
Zero during standby; +5 VDC when
Injecting
injecting; immediately returns to zero
at injection end - no delay.
17 INJECT
Zero during standby; +15 VDC when
Inject
injecting; should stay at +15 VDC for
two seconds after the end of the
injection, then return to zero. With
ECG injections, stays high for one
second at end of injection.
18 A.G.
Reference for other test point grounds.
Analog Ground
19 +15V
+15 +/- 0.5 VDC.
20 -15V
-15 +/- 0.5 VDC.
21 +5V
+5 +/- 0.25 VDC
.
22 MVOLTS
Injecting, 0 - +90 VDC, depending on
Motor Volts
speed. During forward load, typically
measures 7-15 VDC. Reverse load -
less than 0 VDC.
Содержание ProVis Mark V
Страница 1: ...Service Manual KMP 870 98411 T 205 Rev D...
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 15: ...Operating Guide 3 1 3 Operating Guide...
Страница 16: ...Mark V ProVis Injection System 3 2...
Страница 30: ...Mark V ProVis Injection System 7 4 Figure 7 1 Block Diagram Servo Control Card...
Страница 31: ...Servo Control Card 7 5...
Страница 37: ...Mark V ProVis Injection System 9 4 Figure 9 1 Block Diagram Control Panel Digital Logic...
Страница 39: ...Mark V ProVis Injection System 9 6 Figure 9 2 Block Diagram Control Panel DC to DC Converter...
Страница 49: ...Mark V ProVis Injection System 10 10 NOTES...
Страница 60: ...Injector Head 13 7 Figure 13 1 Block Diagram Head Card...
Страница 61: ...Mark V ProVis Injection System 13 8 NOTES...
Страница 64: ...Imaging System Interface 14 3 Figure 14 1 Connector Locations...
Страница 67: ...Mark V ProVis Injection System 14 6 Figure 14 2 Block Diagram Imaging System Interface...
Страница 71: ...Mark V ProVis Injection System 14 10 NOTES...
Страница 107: ...Mark V ProVis Injection System 15 36 NOTES...