Mark V ProVis Injection System
2 - 4
WARNINGS
Dangerous voltages exist within the injection system that can shock,
burn, or cause deat
h. To avoid injury, the system should be opened and
serviced by qualified service personnel only. Disconnect the system from
line power before cleaning or attempting to perform any maintenance or
repairs.
Mains Voltage Hazard.
To avoid dangerous voltages, do not remove con-
nector J130 while power is applied to the system. Always disconnect the
system from line power before removing connector J130
Live Voltage Hazard.
Avoid contact with J40 pins. Ensure that the connec-
tor cover is in place or cable is connected.
Possible explosion hazard in the presence of flammable anesthetic
gases.
The injection
system is not designed for use in association with
anesthetic gases and equipment.
Electronic assemblies contain potentially hazardous materials.
Dis-
pose of system components or accessories properly. Follow all local regu-
lations for the recycling or disposal of electronic assemblies, or contact
MEDRAD Service for assistance.
Worn power cords or control cables may shock, injure, or cause
death
. Examine power cords and cables for cuts, frays, or any other visible
damage. Do not use the system if any of the cords or cables show signs of
damage. Any damaged or worn connection cables or power cords should
be replaced.
Check for proper voltage and frequency before connecting the injec-
tor to an electrical outlet.
Failure to do so may result in personal injury or
equipment damage/malfunction. Check the voltage and frequency marked
on the back of the unit. Ensure that the outlet providing power to the injector
supplies a voltage, frequency, and volt-ampere rating within the range
specified.
Do not use an extension cord or power cord adaptor.
Connect the
injector directly into a properly grounded AC outlet.
Since the power cord
supplies a safety ground to the unit, using an extension cord will compro-
mise ground quality and the injector could become unsafe.
Do not immerse any injector components in water or any type of
cleaning solution.
Fluid entry may result in a shock hazard.
Ensure that FluiDot labels are clearly visible on all pressure jackets.
FluiDot labels are intended to help in the avoidance of air embolization,
which could result in patient injury or death.
Injury may result from springing action when injector head is
removed from the arm.
Move the arm to the maximum upright position
before removing the injector head. When re-mounting the injector head,
ensure that the retention knob is secured to prevent the injector head from
falling from the arm.
!
!
Содержание ProVis Mark V
Страница 1: ...Service Manual KMP 870 98411 T 205 Rev D...
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 15: ...Operating Guide 3 1 3 Operating Guide...
Страница 16: ...Mark V ProVis Injection System 3 2...
Страница 30: ...Mark V ProVis Injection System 7 4 Figure 7 1 Block Diagram Servo Control Card...
Страница 31: ...Servo Control Card 7 5...
Страница 37: ...Mark V ProVis Injection System 9 4 Figure 9 1 Block Diagram Control Panel Digital Logic...
Страница 39: ...Mark V ProVis Injection System 9 6 Figure 9 2 Block Diagram Control Panel DC to DC Converter...
Страница 49: ...Mark V ProVis Injection System 10 10 NOTES...
Страница 60: ...Injector Head 13 7 Figure 13 1 Block Diagram Head Card...
Страница 61: ...Mark V ProVis Injection System 13 8 NOTES...
Страница 64: ...Imaging System Interface 14 3 Figure 14 1 Connector Locations...
Страница 67: ...Mark V ProVis Injection System 14 6 Figure 14 2 Block Diagram Imaging System Interface...
Страница 71: ...Mark V ProVis Injection System 14 10 NOTES...
Страница 107: ...Mark V ProVis Injection System 15 36 NOTES...