Mark V ProVis Injection System
2 - 6
Do not apply voltage to ISI connector J40 pins 5,6,7,8 or 9.
Equip-
ment damage or malfunction may result.
Do not move the injector or pedestal by pulling on the syringe
heater cable.
Equipment damage or malfunction may result.
Do not
connect injector head extension cables in parallel.
Connecting the
extension cables in parallel, or extending the total extension length
beyond 100 feet (30.5 m), can adversely effect injector performance and
specifications.
NOTE:
All relevant institutional, local, or national safety regula-
tions related to cable routing and installation should be
followed.
Do not apply voltage to external start lines if the injector is being
started by an external start connection.
Provide only a switch closure.
If the external circuit contains excessive line frequency noise or voltage
transients, injector damage or malfunction may result.
The Remote Start Signal from the film changer programmer (at pins
7 and 9 of J40), must never contain voltage or AC noise.
Any voltage
applied to these pins can damage the injector and void the warranty.
Excess noise on these lines may cause intermittent problems of prema-
ture, or no injections.
The Remote Start Signal from the film changer programmer must
remain closed throughout the duration of the injection.
If the start
signal is removed before the injection volume limits when armed in the
SINGLE mode, the injection will stop, the unit will disarm, and display
“PREMATURE TERMINATION” on the Control Panel. If the start signal
is removed before the injection volume limits when armed in the MULTI-
ple mode, the injection will stop, the Mechanical Stop will reposition at
the selected volume, and the unit will remain armed. If a start signal is
received at this time, an additional injection will result at the selected
Flow Rate and Volume.
When starting the injection process from a remote location by the
use of a relay closure in the film changer programmer, always
install a “panic button” in series with the start signal.
This function
must be provided in the event that the operator must immediately termi-
nate the injection. This button should be installed in a convenient loca-
tion, properly labeled, and instructions for use provided to all users of the
injection system.
Содержание ProVis Mark V
Страница 1: ...Service Manual KMP 870 98411 T 205 Rev D...
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 15: ...Operating Guide 3 1 3 Operating Guide...
Страница 16: ...Mark V ProVis Injection System 3 2...
Страница 30: ...Mark V ProVis Injection System 7 4 Figure 7 1 Block Diagram Servo Control Card...
Страница 31: ...Servo Control Card 7 5...
Страница 37: ...Mark V ProVis Injection System 9 4 Figure 9 1 Block Diagram Control Panel Digital Logic...
Страница 39: ...Mark V ProVis Injection System 9 6 Figure 9 2 Block Diagram Control Panel DC to DC Converter...
Страница 49: ...Mark V ProVis Injection System 10 10 NOTES...
Страница 60: ...Injector Head 13 7 Figure 13 1 Block Diagram Head Card...
Страница 61: ...Mark V ProVis Injection System 13 8 NOTES...
Страница 64: ...Imaging System Interface 14 3 Figure 14 1 Connector Locations...
Страница 67: ...Mark V ProVis Injection System 14 6 Figure 14 2 Block Diagram Imaging System Interface...
Страница 71: ...Mark V ProVis Injection System 14 10 NOTES...
Страница 107: ...Mark V ProVis Injection System 15 36 NOTES...