Mark V ProVis Injection System
16 - 32
Power Switch
Power Switch
Removal
1. Remove the injector from the power source.
2. Remove the power switch by prying gently with a screwdriver.
3. Disconnect the wires from the switch.
Power Switch
Replacement
1 Connect the power switch wires as shown below.
Figure 16.10: Power Switch Terminal Orientations
2. Place the switch in the unit without inserting fully. Connect the
injector to a power source, then apply power.
3. If the switch is oriented correctly, the On/Off label will agree with
the position of the switch. If this is not the case, remove and
reverse the switch. When oriented correctly, fully insert the switch
into the case.
Содержание ProVis Mark V
Страница 1: ...Service Manual KMP 870 98411 T 205 Rev D...
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 15: ...Operating Guide 3 1 3 Operating Guide...
Страница 16: ...Mark V ProVis Injection System 3 2...
Страница 30: ...Mark V ProVis Injection System 7 4 Figure 7 1 Block Diagram Servo Control Card...
Страница 31: ...Servo Control Card 7 5...
Страница 37: ...Mark V ProVis Injection System 9 4 Figure 9 1 Block Diagram Control Panel Digital Logic...
Страница 39: ...Mark V ProVis Injection System 9 6 Figure 9 2 Block Diagram Control Panel DC to DC Converter...
Страница 49: ...Mark V ProVis Injection System 10 10 NOTES...
Страница 60: ...Injector Head 13 7 Figure 13 1 Block Diagram Head Card...
Страница 61: ...Mark V ProVis Injection System 13 8 NOTES...
Страница 64: ...Imaging System Interface 14 3 Figure 14 1 Connector Locations...
Страница 67: ...Mark V ProVis Injection System 14 6 Figure 14 2 Block Diagram Imaging System Interface...
Страница 71: ...Mark V ProVis Injection System 14 10 NOTES...
Страница 107: ...Mark V ProVis Injection System 15 36 NOTES...