Troubleshooting Guide
15 - 5
Test Point / Description
Typical Values
23 PLNG. MOT. VOLTS LOW
Less than 1 VDC while injecting
Plunger Motor VDC Low Side
or forward load. +1 VDC with
1200 PSI. Reverse load 7-15
VDC.
24 FLOW SELECTED
Measure in ml/sec - 60 ml
Flow Selected
syringe, +0.38 VDC/ml. 150 ml
syringe, +0.16 VDC/ml (+6.4
VDC at 40 ml/sec). 200 ml
syringe, +0.12 VDC (+4.8 VDC
at 40 ml/sec). In ml/min or ml/hr,
less than 0.2 VDC.
25 150/200 ENABLE
60 ml = 5 VDC
150/200 Enable
150 ml = 0 VDC
200 ml = 5 VDC
26 F/C START
F/C Relay Open = 26 VDC
Film Changer Start
F/C Relay Closed = 0 VDC
27 ARM LT. ENABLE
Light off = 0 VDC
Arm Light Enable
Light on = 5 VDC
Flashing = Pulses 0 to 5 VDC
28 SYSTEM OVERRIDE
Standby = 0 VDC
System Override
Active state = 5 VDC
29 AUX MONITOR
Standby = -14 VDC
Aux. Monitor
Active state = 13 VDC
30 +26V
+26
VDC
Содержание ProVis Mark V
Страница 1: ...Service Manual KMP 870 98411 T 205 Rev D...
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 15: ...Operating Guide 3 1 3 Operating Guide...
Страница 16: ...Mark V ProVis Injection System 3 2...
Страница 30: ...Mark V ProVis Injection System 7 4 Figure 7 1 Block Diagram Servo Control Card...
Страница 31: ...Servo Control Card 7 5...
Страница 37: ...Mark V ProVis Injection System 9 4 Figure 9 1 Block Diagram Control Panel Digital Logic...
Страница 39: ...Mark V ProVis Injection System 9 6 Figure 9 2 Block Diagram Control Panel DC to DC Converter...
Страница 49: ...Mark V ProVis Injection System 10 10 NOTES...
Страница 60: ...Injector Head 13 7 Figure 13 1 Block Diagram Head Card...
Страница 61: ...Mark V ProVis Injection System 13 8 NOTES...
Страница 64: ...Imaging System Interface 14 3 Figure 14 1 Connector Locations...
Страница 67: ...Mark V ProVis Injection System 14 6 Figure 14 2 Block Diagram Imaging System Interface...
Страница 71: ...Mark V ProVis Injection System 14 10 NOTES...
Страница 107: ...Mark V ProVis Injection System 15 36 NOTES...