Mark ProVis Injection System
4 - 4
Injector Head
Articulating Arm
The injector head support arm is typically associated with pedestal type
injection systems and free standing injector head stands. Carefully
inspect the areas shown below for any signs of damage.
1. Inspect all parts of the mounting system for cracks and other defects
that would weaken the assembly.
2. Ensure that the mounting system is securely assembled, with no
loose parts. The system should be stable with the head installed.
3. Ensure that the system moves smoothly in all directions, with no bind-
ing, scraping, or drifting.
4. Verify that all cabling is tied back and does not interfere with the
movement of the supporting parts or the injector head.
NOTE
: The injector head pivot knuckle should rest flat and rotate
smoothly on the support arm.
NOTE
: The head cable can wrap around the support arm, causing
damage to the head cable, and possibly lifting the injector
head from the support arm. Contact MEDRAD Factory
Service for support arm replacement information.
Counterpoise
Systems
1. Inspect all parts of the mounting system for cracks and other defects
that would weaken the assembly.
2. Ensure that the mounting system is securely assembled, with no
loose parts. The system should be stable with the head installed.
3. Ensure that the system moves smoothly in all directions, with no bind-
ing, scraping, or drifting.
4. Verify that all cabling is tied back and does not interfere with the
movement of the supporting parts or the injector head.
NOTE
: All relevant guidelines for institutional, local, or national
safety recommendations related to cable routing and
installation should be followed.
NOTE
: Rotating the counterpoise consistently in one direction can
cause the head extension cable to wrap around the coun-
terpoise, causing possible equipment damage.
NOTE
: The head cable can wrap around the support arm, causing
damage to the head cable, and possibly lifting the injector
head from the support arm. Contact MEDRAD Factory
Service for support arm replacement information.
Содержание ProVis Mark V
Страница 1: ...Service Manual KMP 870 98411 T 205 Rev D...
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 15: ...Operating Guide 3 1 3 Operating Guide...
Страница 16: ...Mark V ProVis Injection System 3 2...
Страница 30: ...Mark V ProVis Injection System 7 4 Figure 7 1 Block Diagram Servo Control Card...
Страница 31: ...Servo Control Card 7 5...
Страница 37: ...Mark V ProVis Injection System 9 4 Figure 9 1 Block Diagram Control Panel Digital Logic...
Страница 39: ...Mark V ProVis Injection System 9 6 Figure 9 2 Block Diagram Control Panel DC to DC Converter...
Страница 49: ...Mark V ProVis Injection System 10 10 NOTES...
Страница 60: ...Injector Head 13 7 Figure 13 1 Block Diagram Head Card...
Страница 61: ...Mark V ProVis Injection System 13 8 NOTES...
Страница 64: ...Imaging System Interface 14 3 Figure 14 1 Connector Locations...
Страница 67: ...Mark V ProVis Injection System 14 6 Figure 14 2 Block Diagram Imaging System Interface...
Страница 71: ...Mark V ProVis Injection System 14 10 NOTES...
Страница 107: ...Mark V ProVis Injection System 15 36 NOTES...