Mark V ProVis Injection System
16 - 18
Injector Head
- Plunger Position Pot
(continued)
8. Arm the injector. When the Mechanical Stop is in position, mea-
sure the wiper voltage on the white/yellow wire of the mechanical
stop pot. The voltage must be 0.860 VDC +/-0.020 VDC. If voltage
is within specification, skip to step 14.
9. Disconnect one of the mechanical stop motor leads. Manually
move the Mechanical Stop plate forward approximately 1/8 inch
(3.2 mm).
10. Advance the plunger until the plunger position pot wiper voltage
(terminal 2) is 1.010 VDC.
11. Manually reverse the Mechanical Stop plate into the plunger plate
until the mechanical stop microswitch opens (noted by an audible
click). The DVM can be used on the microswitch to detect when
the microswitch opens.
12. Disengage the Mechanical Stop pot and adjust for 0.860 VDC.
13. Reengage the pot gear, and tighten the pot bracket screws.
Reverse plunger to the rear limit. Disconnect DVM from head,
then reconnect the Mechanical Stop motor lead.
14. Remove injector power. Ensure that all pot screws are secure,
then install the cable bracket on the rear casting.
15. If required, attach the metal clamp and head cable on the cable
bracket.
16. Secure the wires and head cable with cable ties.
17. Reassemble the injector head.
18. Ensure that the proper syringe size is selected (look at the indica-
tors on the injector head). If the incorrect syringe size is selected,
press RECALL, 50, RECALL, then select the proper syringe size.
Содержание ProVis Mark V
Страница 1: ...Service Manual KMP 870 98411 T 205 Rev D...
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 15: ...Operating Guide 3 1 3 Operating Guide...
Страница 16: ...Mark V ProVis Injection System 3 2...
Страница 30: ...Mark V ProVis Injection System 7 4 Figure 7 1 Block Diagram Servo Control Card...
Страница 31: ...Servo Control Card 7 5...
Страница 37: ...Mark V ProVis Injection System 9 4 Figure 9 1 Block Diagram Control Panel Digital Logic...
Страница 39: ...Mark V ProVis Injection System 9 6 Figure 9 2 Block Diagram Control Panel DC to DC Converter...
Страница 49: ...Mark V ProVis Injection System 10 10 NOTES...
Страница 60: ...Injector Head 13 7 Figure 13 1 Block Diagram Head Card...
Страница 61: ...Mark V ProVis Injection System 13 8 NOTES...
Страница 64: ...Imaging System Interface 14 3 Figure 14 1 Connector Locations...
Страница 67: ...Mark V ProVis Injection System 14 6 Figure 14 2 Block Diagram Imaging System Interface...
Страница 71: ...Mark V ProVis Injection System 14 10 NOTES...
Страница 107: ...Mark V ProVis Injection System 15 36 NOTES...