Mark V ProVis Injection System
15 - 16
ENTER THREE CHARACTER PASSWORD
Program 65 has been recalled to change the programmed head con-
figuration. Enter the service access password, then select the proper
head configuration.
ENTER TITLE OR STORE
STORE sequence prompt that is displayed after a PPI number has
been entered; allows the user to add a title to the stored injection.
Procedure: If a title is desired, enter numbers on the keypad or press
SHIFT to enter letters. If no title is desired, press STORE again.
ENTER VALUES OR ARMSIZE XXX/TVOL XXX
The unit is in a stand-by condition. The operator may enter new val-
ues, or keep the existing values shown or arm by pressing one of the
ARM buttons. The alternating message on the right displays the pro-
grammed syringe size and the total Volume selected.
ENTER VOLUME
The Delta 4 button was pressed. The injector is now ready for a Vol-
ume value to be entered.
ESTABLISHING COMMUNICATIONS
This message is displayed on the Control Panel after power-up or
recovery from a system monitor condition. This message implies that
the Control Panel is attempting to communicate with the Main Unit,
not that a system monitor condition exists. If an error occurs, another
message will be prompted to specify the nature of the communica-
tions failure.
**EXT START SHORTED
There is a short in the external start circuit, or the external start switch
is closed during the arming procedure.
Procedure:
1. Ensure that the start button is not pressed.
2. Disconnect any auxiliary start connections (footswitch,
auxiliary handswitch, etc.), then retry.
3. If the injector now functions properly using an external start
switch, check the start switch and cord reel for shorts and
improper closure.
4. If the condition remains with the cord reel disconnected,
replace the I/O Card.
EXTERNAL DISARM SIGNAL ACTIVE
External disarm input, available through the Imaging System Interface
ISI-100, (J40) is active. The injector cannot be armed until this input is
open.
Содержание ProVis Mark V
Страница 1: ...Service Manual KMP 870 98411 T 205 Rev D...
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 15: ...Operating Guide 3 1 3 Operating Guide...
Страница 16: ...Mark V ProVis Injection System 3 2...
Страница 30: ...Mark V ProVis Injection System 7 4 Figure 7 1 Block Diagram Servo Control Card...
Страница 31: ...Servo Control Card 7 5...
Страница 37: ...Mark V ProVis Injection System 9 4 Figure 9 1 Block Diagram Control Panel Digital Logic...
Страница 39: ...Mark V ProVis Injection System 9 6 Figure 9 2 Block Diagram Control Panel DC to DC Converter...
Страница 49: ...Mark V ProVis Injection System 10 10 NOTES...
Страница 60: ...Injector Head 13 7 Figure 13 1 Block Diagram Head Card...
Страница 61: ...Mark V ProVis Injection System 13 8 NOTES...
Страница 64: ...Imaging System Interface 14 3 Figure 14 1 Connector Locations...
Страница 67: ...Mark V ProVis Injection System 14 6 Figure 14 2 Block Diagram Imaging System Interface...
Страница 71: ...Mark V ProVis Injection System 14 10 NOTES...
Страница 107: ...Mark V ProVis Injection System 15 36 NOTES...