Troubleshooting Guide
15 - 31
150/200 SYRINGE SYSTEM INSTALLED
The user attempted a size change to a 60 ml syringe from the Recall
50 utility program when the system software/hardware is configured
for the 150/200 syringe system. Use Recall 65 to access the 60/150
syringe system and select the 60 ml syringe size.
NOTE
: The
Mark V ProVis
injector should not be configured
for 60 ml syringe operation.
SERVICE UTILITY MESSAGES
The following messages are displayed when service utility programs
are accessed. Under normal circumstances, an operator of the injec-
tor will never see these messages displayed.
CHECK HANDSHAKE LINES
CHECK XMT/RCV LINES
CHECK HANDSHAKE LINES
MOVE PLUNGER TO FULL REVERSE
TESTING INTERFACE - XMIT/RCV=
Содержание ProVis Mark V
Страница 1: ...Service Manual KMP 870 98411 T 205 Rev D...
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 15: ...Operating Guide 3 1 3 Operating Guide...
Страница 16: ...Mark V ProVis Injection System 3 2...
Страница 30: ...Mark V ProVis Injection System 7 4 Figure 7 1 Block Diagram Servo Control Card...
Страница 31: ...Servo Control Card 7 5...
Страница 37: ...Mark V ProVis Injection System 9 4 Figure 9 1 Block Diagram Control Panel Digital Logic...
Страница 39: ...Mark V ProVis Injection System 9 6 Figure 9 2 Block Diagram Control Panel DC to DC Converter...
Страница 49: ...Mark V ProVis Injection System 10 10 NOTES...
Страница 60: ...Injector Head 13 7 Figure 13 1 Block Diagram Head Card...
Страница 61: ...Mark V ProVis Injection System 13 8 NOTES...
Страница 64: ...Imaging System Interface 14 3 Figure 14 1 Connector Locations...
Страница 67: ...Mark V ProVis Injection System 14 6 Figure 14 2 Block Diagram Imaging System Interface...
Страница 71: ...Mark V ProVis Injection System 14 10 NOTES...
Страница 107: ...Mark V ProVis Injection System 15 36 NOTES...