Disassembly and Reassembly
16 - 27
Injector Head
- Control Panel
NOTE
: Removal and replacement of the injector head Control
Panel requires major disassembly of the injector head.
Although the procedures are included here, MEDRAD
does not recommended field repair of these items, and
suggests that the injector head be returned to MEDRAD
Service for these repairs.
Control Panel
Removal
1. After disassembly (see Page 16 - 13), place the Injector Head top-
side up.
2. Remove the four flat-head screws mounting the Control Panel to the
top of the Injector Head.
3. Remove the screw and nut securing the arm lamps to the plate.
4. Remove the head cable connectors, then carefully remove the Con-
trol Panel and attached board.
5. Remove the screws mounting the Head Card and card shield to the
plate. Disconnect the ribbon cable at the connector.
Control Panel
Replacement
1. Reconnect the ribbon cable from the push-button switch to the PC
card.
2. Place the top plate and card on a smooth flat surface, top side down.
3. Gently tilt the card upward, placing the card (component side down)
on the top plate.
NOTE:
Use extreme caution to avoid creasing of the ribbon
cable.
4. Mount the PC card and card shield to the top plate with screws previ-
ously removed, ensuring that the ribbon cable is properly positioned,
not crimped or pinched.
5. Position the Top Plate on the head, then mount with the four screws
previously removed.
6. Attach Head Cable connectors and arm lamps.
7. Connect the Injector Head Cable to the rear connector panel of the
injector, then apply power to the system. Ensure that the plunger
moves freely in both the forward and reverse directions.
8. Arm the injector. Verify that the head will arm properly.
9. If the Injector Head is functioning properly, reassemble following the
procedure on Page 16 - 13.
Содержание ProVis Mark V
Страница 1: ...Service Manual KMP 870 98411 T 205 Rev D...
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 15: ...Operating Guide 3 1 3 Operating Guide...
Страница 16: ...Mark V ProVis Injection System 3 2...
Страница 30: ...Mark V ProVis Injection System 7 4 Figure 7 1 Block Diagram Servo Control Card...
Страница 31: ...Servo Control Card 7 5...
Страница 37: ...Mark V ProVis Injection System 9 4 Figure 9 1 Block Diagram Control Panel Digital Logic...
Страница 39: ...Mark V ProVis Injection System 9 6 Figure 9 2 Block Diagram Control Panel DC to DC Converter...
Страница 49: ...Mark V ProVis Injection System 10 10 NOTES...
Страница 60: ...Injector Head 13 7 Figure 13 1 Block Diagram Head Card...
Страница 61: ...Mark V ProVis Injection System 13 8 NOTES...
Страница 64: ...Imaging System Interface 14 3 Figure 14 1 Connector Locations...
Страница 67: ...Mark V ProVis Injection System 14 6 Figure 14 2 Block Diagram Imaging System Interface...
Страница 71: ...Mark V ProVis Injection System 14 10 NOTES...
Страница 107: ...Mark V ProVis Injection System 15 36 NOTES...