Q-Sense Technical Reference Manual
Proprietary
168 of 188
№ Error Message
Solution
19.
Time mark missed. Error Code: 400D
An error has occurred in the Q-SENSE
processor during run time.
1.
Click OK and continue working as
usual.
2.
Restart the Q-SENSE system and
software.
If the problem persists, send the
corresponding error code to the Medoc
service team.
20.
Sample buffer full. Error Code: 2001
An error occurred in the Q-SENSE
processor during run time.
1.
Click OK and continue working as
usual.
If the problem persists, send the
corresponding error code to the Medoc
service team.
21.
Finite ramp by temperature timeout.
Thermode temperature: 35 (°C). Error
Code: 2002
The Q-SENSE system was unable to
complete the requested command in the
allowed time, and has therefore switched
to Safe Mode in order to prevent any
damage.
1.
Check that the Fan mode is according
to requirement.
2.
Run Self Test to verify that the Q-
SENSE system is enabled to continue
work.
If the problem persists, stop using the Q-
SENSE system and contact Medoc
service.
22.
Safety cold duration limit violation.
Temperature level: 20 (°C), Time entry:
10(sec). Error Code: 800E
The Q-SENSE system has detected that
the safety limitation was violated, and
therefore switched to Safe Mode in order
to prevent any damage.
1.
Make sure that the program you are
trying to run complies with the safety
limitations.
2.
Run Self Test to verify that the Q-
SENSE system is enabled to continue
work.
23.
System could not establish communication!
Retry: The system will retry to establish
communication.
Cancel: Continue working without
connecting to device.
The program was unable to establish
communication with the Q-SENSE
system.
Select one of the presented options and
continue.
Содержание TSA 2001
Страница 1: ...Proprietary 1 of 188 Small Fiber Test Technical Reference Manual ...
Страница 35: ...Q Sense Technical Reference Manual Proprietary 35 of 188 Figure 7 Q SENSE System Wiring Schema ...
Страница 99: ...Q Sense Technical Reference Manual Proprietary 99 of 188 Analysis Report Examples ...
Страница 109: ...Q Sense Technical Reference Manual Proprietary 109 of 188 ...
Страница 181: ...Q Sense Technical Reference Manual Proprietary 181 of 188 Figure 114 Pain Rating Limits ...