Q-Sense Technical Reference Manual
Proprietary
64 of 188
Parameter
Description
Update on
The date on which the latest changes were made and saved.
Start test/End test
Defines if these events are indicated by sound.
Table 12: Program Parameters
Parameter
Description
Modality
Four different types of sensation can be delivered to the
subject:
Warm sensation; Cold sensation; Hot pain.
Baseline (°C)
Temperature from which each new stimulation sequence begins
– between 28°C to 40°C for Q-sense thermode.
Rate (°C/sec)
Defines the temperature rate of increase per second, from the
baseline until the subject’s response.
Type an acceptable value: 0.1 to 2°C/Sec for Q-sense
thermode
Return Rate
(°C/sec)
Defines the temperature rate of decrease per second, from the
temperature at which the subject responded, down to the
baseline.
Type an acceptable value: .. 0.1-1°C/sec for Q-sense thermode.
Number of Trials
The number of trials in the sequence – between 1 to 100.
Randomize with next Randomizes the next trial or sequence.
ISI Min (sec) /
ISI Max (sec)
Variable time between trials. Type acceptable value: 2-100.
Standard Events
Enables the system to alert you when stimulation procedure
arrives at some standard events.
Содержание TSA 2001
Страница 1: ...Proprietary 1 of 188 Small Fiber Test Technical Reference Manual ...
Страница 35: ...Q Sense Technical Reference Manual Proprietary 35 of 188 Figure 7 Q SENSE System Wiring Schema ...
Страница 99: ...Q Sense Technical Reference Manual Proprietary 99 of 188 Analysis Report Examples ...
Страница 109: ...Q Sense Technical Reference Manual Proprietary 109 of 188 ...
Страница 181: ...Q Sense Technical Reference Manual Proprietary 181 of 188 Figure 114 Pain Rating Limits ...