Q-Sense Technical Reference Manual
Proprietary
149 of 188
17
Appendix E – Preventive and Periodical
Maintenance
This chapter describes periodic maintenance procedures.
17.1
Periodical System Run and Visual Check
A periodical system run and visual examination are necessary for maintaining performance of Q-
SENSE system. The periodical run keeps all Q-SENSE accessories and subunits in operational
condition at all times. Periodical system run is required, if the system or the supplied Thermode is
not used on a regular basis, or if it has not been in use for a period of one month or more,
Visual check is required in order to make sure that there is no
physical damage to system components, particularly to the
Thermodes.
1.
Check all accessories and Thermodes visually.
1. Check that accessories connector is intact.
2. Check that accessories body is intact.
3. Check that there is no visible damage to the Thermode heads.
2.
Install all available Thermode types (Q-sense type), according to your license type, to the Q-
SENSE system.
3.
Connect all available Q-Sense accessories (Response Unit) to the Q-SENSE system.
4.
Connect the communication cable from the Q-Sense system to the computer running MEDOC
Main Station.
5.
Power ON the Q-SENSE system.
6.
Start MEDOC Main Station and ensure that communication is established.
7.
Run the system for about 10 minutes; check for any visible malfunctions.
8.
Run a few different test protocols such as Limits.
9.
Repeat steps 8 and 9 with all the Thermodes supplied with the system.
Switch Thermodes only when the Q-Sense system is turned OFF!
Remember to change calibration tables prior switching to different
Q-sense Thermode type (30x30mm).
10.
If the system is not planned for use for three weeks or more, perform the following
instructions:
a.
Turn OFF the Q-SENSE system through the main switch.
b.
Disconnect all accessories from the Q-SENSE system, and store them in a dry place.
c.
Disconnect all Thermodes from the Q-SENSE system, cover the Thermode heads with
soft cloth and store in dark dry place.
Содержание TSA 2001
Страница 1: ...Proprietary 1 of 188 Small Fiber Test Technical Reference Manual ...
Страница 35: ...Q Sense Technical Reference Manual Proprietary 35 of 188 Figure 7 Q SENSE System Wiring Schema ...
Страница 99: ...Q Sense Technical Reference Manual Proprietary 99 of 188 Analysis Report Examples ...
Страница 109: ...Q Sense Technical Reference Manual Proprietary 109 of 188 ...
Страница 181: ...Q Sense Technical Reference Manual Proprietary 181 of 188 Figure 114 Pain Rating Limits ...