Q-Sense Technical Reference Manual
Proprietary
17 of 188
Adverse Reaction: Skin irritation (in addition to pain sensation) beneath the probe has
been reported with the use of a stimulator, which was based on similar technology as
the Q-Sense device.
Be aware of potential risk of skin damage caused by wrong parameter
combination.
Use of this equipment adjacent to or stacked with other equipment should be
avoided because it could result in improper operation. If such use is necessary,
this equipment and the other equipment should be observed to verify that they
are operating normally.
of accessories, transducers and cables other than those specified or provided by
the manufacturer of this equipment could result in increased electromagnetic
emissions or decreased electromagnetic immunity of this equipment and result
in improper operation.
Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches)
to any part of the [
ME
EQUIPMENT
or
ME
SYSTEM
], including cables specified by the
manufacturer. Otherwise, degradation of the performance of this equipment
could result.
NOTE The EMISSIONS characteristics of this equipment make it suitable for use
in industrial areas and hospitals (CISPR 11 class A). If it is used in a residential
environment (for which CISPR 11 class B is normally required) this equipment
might not offer adequate protection to radio-frequency communication services.
The user might need to take mitigation measures, such as relocating or re-
orienting the equipment
1.5.2
Cautions
Proper use of this device depends on careful reading of all instructions and labels.
Turn OFF system power before connecting or disconnecting any system component(s)
or accessories. Otherwise, you may damage the device(s).
The Thermode is very delicate and can easily be damaged. Therefore, handle with
care.
If you disconnect any cables, take care to reconnect them correctly to prevent damage
to the system or components.
Inspect the power cord often for fraying or other damage. DO NOT operate the
apparatus if the power cord or plug is damaged.
The computer that is used to operate the Q-SENSE system must not be connected to a
network while it is used for running tests.
The Q-SENSE system does not require special precautions regarding EMC, and needs
to be installed and put into service according to this manual.
Portable and mobile RF communications equipment can affect medical electrical
equipment.
Use caution when using the Q-SENSE Thermode on patients with suspected
neuropathies as they may be more susceptible to soft tissue or nerve damage at
extreme temperatures. Also, patients with neuropathies may not be able to
properly discontinue use of the device during prolonged hot or cold stimulation.
Содержание TSA 2001
Страница 1: ...Proprietary 1 of 188 Small Fiber Test Technical Reference Manual ...
Страница 35: ...Q Sense Technical Reference Manual Proprietary 35 of 188 Figure 7 Q SENSE System Wiring Schema ...
Страница 99: ...Q Sense Technical Reference Manual Proprietary 99 of 188 Analysis Report Examples ...
Страница 109: ...Q Sense Technical Reference Manual Proprietary 109 of 188 ...
Страница 181: ...Q Sense Technical Reference Manual Proprietary 181 of 188 Figure 114 Pain Rating Limits ...