Q-Sense Technical Reference Manual
Proprietary
25 of 188
1.10
Technical Data
Medoc reserves the right to change specifications without prior notice,
in line with the company policy of constant product improvement
The table below describes the technical specifications and capabilities of the Q-Sense system.
Table 2: Q-SENSE Specifications
Parameter
Specification
Thermode Active Area
30x30 mm
Thermode temperature Range
20 – 50 °C
Heating Rate
0 – 2 °C/Sec ± 10%
Cooling Rate
0 - 1 °C/Sec ± 10%
Stimulation Methods
Limits, Levels & TSL
Ramp & Hold (optional)
Set-Point Resolution
0.1 °C
Display Resolution
0.1 °C
Repeatability
± 0.3 °C
Absolute Accuracy
±0.3 °C
Operating Mode
Intervals*
*(Short pause between patients may be
required)
Minimum Inter Stimuli Interval (ISI)
2 sec
Maximum Consecutive Stimuli
10
Baseline Thermode Temperature Range
28 – 40 °C
Programmable Parameters
Method, Adaptation Temperature, Heating Rate,
Cooling Rate, Number of Stimuli, ISI, Sound
Option, Randomize Option
Communication with PC
USB
Communication with Printer
Via PC
Standards
IEC 60601-1; IEC 60601-1-2
Approvals
CE, FDA
Operating Voltage
12 VDC, 5A
Power Supply
100 – 240 VAC, 4 – 2 A, 50/60 Hz
Power Consumption
Approx. 60W
Dimensions
31cm X 28 cm X 16cm (table or wall mount)
Shipping: 53cm X 34cm X 22cm
Weight
Approx. 4.5 Kg
Shipping: Approx. 10 Kg
Software
Propriety Software Application (Included)
Computer Requirements (PC not
included)
Notebook PC (Atom CPU not supported)
At least 2 GB RAM, 6 GB Free disc space.
Windows 7/8 (32 & 64 bit), Windows XP (32bit)
sp. 3 and above.
Содержание TSA 2001
Страница 1: ...Proprietary 1 of 188 Small Fiber Test Technical Reference Manual ...
Страница 35: ...Q Sense Technical Reference Manual Proprietary 35 of 188 Figure 7 Q SENSE System Wiring Schema ...
Страница 99: ...Q Sense Technical Reference Manual Proprietary 99 of 188 Analysis Report Examples ...
Страница 109: ...Q Sense Technical Reference Manual Proprietary 109 of 188 ...
Страница 181: ...Q Sense Technical Reference Manual Proprietary 181 of 188 Figure 114 Pain Rating Limits ...