Q-Sense Technical Reference Manual
Proprietary
169 of 188
№ Error Message
Solution
24.
The thermometer does not respond.
1. Check that the thermometer is turned
on.
2. Check that the thermometer is
connected to the correct COM (see setting
> software setting > calibration COM port).
Automatic calibration had failed, because
the program could not communicate with
the thermometer.
Follow the displayed instructions, and
retry.
25.
The measured temperature 51(˚C) is out
of permitted range [20…50].
During automatic calibration the
measured temperature exceeded the
allowed range.
1.
Reduce calibration range (increase
minimum and decrease maximum)
through Automatic calibration
properties.
2.
Retry calibrating.
If calibration is still out of range,
calibrate the Thermode manually to
bring it back into range.
26.
Have not been able to read temperatures
from the Thermometer for 60 seconds.
1. Verify that the thermometer
communication cables are connected
properly.
2. Verify that the thermometer is turned
on.
3. Restart the automatic calibration
procedure.
The program was unable to communicate
with the thermometer.
Follow the displayed instructions, and
retry.
27.
1. Check that the thermode is properly
placed in the jig and connected to the Q-
Sense.
2. During automatic calibration software
safety protection is lowered; therefore, it is
not allowed to place Thermode on human
subjects
Click YES to continue calibrating.
Click NO to discontinue calibration
Warning notification before automatic
calibration.
Follow the displayed instructions, and
continue.
28.
Can not open thermometer serial
communication port, because it is used by
Medoc Main Station
The selected calibration COM port is in
use by the Q-SENSE system.
1.
Select the appropriate calibration
COM port through Settings
Software
Settings
General.
2.
Retry calibration.
Содержание TSA 2001
Страница 1: ...Proprietary 1 of 188 Small Fiber Test Technical Reference Manual ...
Страница 35: ...Q Sense Technical Reference Manual Proprietary 35 of 188 Figure 7 Q SENSE System Wiring Schema ...
Страница 99: ...Q Sense Technical Reference Manual Proprietary 99 of 188 Analysis Report Examples ...
Страница 109: ...Q Sense Technical Reference Manual Proprietary 109 of 188 ...
Страница 181: ...Q Sense Technical Reference Manual Proprietary 181 of 188 Figure 114 Pain Rating Limits ...