Q-Sense Technical Reference Manual
Proprietary
182 of 188
26
Appendix N – fMRI functionality
26.1
fMRI Thermode
Q-sense fMRI Thermode and accessories are defined and marked as MR
Conditional equipment according to the ASTM F2503–08.
Q-sense MR filters are defined and marked as MR Unsafe equipment
according to the ASTM F2503–08.
WARNING:
Make sure that the Thermode is unblemished in order to prevent potential
harm to the test subject.
WARNING:
To avoid the risk of electric shock, this equipment must only be connected
to a supply mains with protective earth.
The fMRI Q-sense system was tested in the following MR environments:
Field Strength: 3T and 7T; Max. Spatial gradient: 72 mT/m
Max. dB/dt: 346 T/m/s; RF Fields: max. power 35KW pep;
TX bandwidth: 800kHz; RX bandwidth: 500Hz-1MHz; maximal RF field
strength: 24,7 µT CP
SAR: whole body 4 W/kg, head 3.2 W/Kg
The fMRI Q-sense system was tested in the following MR Sequences:
1.
Echo Planer Imaging (EPI) bold (fMRI)
2.
rf_noise and rf_noise spectrum electromagnetic noise interference
(Siemens protocol)
26.1.1
Q-sense system configuration in fMRI mode:
The fMRI Q-sense system includes the following components:
Control room thermode cable
Q-Sense fMRI Thermode
Содержание TSA 2001
Страница 1: ...Proprietary 1 of 188 Small Fiber Test Technical Reference Manual ...
Страница 35: ...Q Sense Technical Reference Manual Proprietary 35 of 188 Figure 7 Q SENSE System Wiring Schema ...
Страница 99: ...Q Sense Technical Reference Manual Proprietary 99 of 188 Analysis Report Examples ...
Страница 109: ...Q Sense Technical Reference Manual Proprietary 109 of 188 ...
Страница 181: ...Q Sense Technical Reference Manual Proprietary 181 of 188 Figure 114 Pain Rating Limits ...