Q-Sense Technical Reference Manual
Proprietary
176 of 188
25.2
CoVAS
Enables continues pain rating using the USB CoVAS device.
25.2.1
CoVAS Calibration
Calibrating the CoVAS ensures the CoVAS range 0 to 100 corresponds with the left most and right
most point of the slider, respectively.
The CoVAS should be calibrated when:
It is used for the first time
Software or firmware are re-installed
To Calibrate the CoVAS:
1.
Connect the CoVAS to the Computer (USB connection).
2.
In the Hardware Settings windows, select the USB CoVAS tab.
Figure 110: CoVAS Calibration Tab
3.
Click the Next button and follow the instructions shown on screen (a short version is given
here)
4.
Move the CoVAS slider all the way to the left and click the Next button.
5.
Move the CoVAS slider all the way to the right and click the Next button.
6.
Click Next to save the calibration results.