QUI-1122
Page 44
Markes International Ltd
T:
+44 (0)1443 230935
F:
+44 (0)1443 231531
www.markes.com
6.5
UNITY–Air Server-xr Electronic Gas Control with the Thermo 1300 Series GC
A UNITY–Air Server-xr can experience electronic pressure control using the Thermo 1300 series GC.
Currently, full ECC control is not available, but the Thermo 1300 Series GC can be modified to supply a
constant pressure or flow to the UNITY–Air Server-xr. The UNITY–Air Server-xr must be configured in the T’d
EPC configuration (see section 3.4.3) and then the
This setup enables carrier gas settings to be stored and recalled as part of the GC(MS) method and allows
constant flow GC methods to be implemented. The solution is to install a standard Thermo Scientific
headspace adapter onto a S/SL inlet (p/n 19050732; TriPlus 300 HS adapter kit for Instant connect SSL
module). The carrier gas supply from this can be used to provide constant flow carrier gas supply to the GC
as described in the note below.
Note:
Changing split conditions on the TD or changing the trap hold time from one run to the next may
change the relative retention times of early eluting compounds.
6.5.1
Installing the Markes TD to the Thermo Scientific TRACE Series 1300 GC
Carrier gas should be supplied via a GAS01 regulator or similar step down regulator to the inlet for the
S/SL module. The carrier gas pressure should be supplied at no more than 10psi above the maximum
pressure required during the GC oven program.
The TD transfer line should pass through an unused injector/detector port or through a modified panel on
the right hand side of the TRACE Series 1300 GC and secured inside the oven. The fused silica can then
be connected to the column via a glass QuickSeal connector or zero dead volume connector.
A standard Thermo Scientific Headspace adapter should be fitted on to the S/SL inlet. Connect a length of
1/16” tubing (shorter is better and stainless steel is preferred to PEEK on GCMS) to the port of the
headspace adapter. Connect the other end of the 1/16” tubing to the carrier gas inlet on the rear of the
TD. The underside (outlet) of the S/SL injector in the GC oven should be fitted with a blanking ferrule or
nut.